Technical Specifications:
Principles :Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye
method, independent Basophil channel
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Parameters 27: WBC, Lym%, Mon%, Neu%, Eos%, Bas%, Lym#, Mon#, Neu#, Eos#, Bas#,
RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT,
LIC%, LIC#, ALY%, ALY#
3 histograms and 1 scatter gram
Sampling Mode :
Autoloader
Closed Tube
Throughput :
Up to 60 samples per hour with autoloader
Up to 50 samples per hour with closed tube
Sample Volume:
Whole Blood: 20 μL
Prediluted: 20 μL
Test Mode:
CBC CBC+DIFF
Performance:
Carryover Precision Linearity:
WBC ≤ 0.5% ≤ 2.0% (4-15 x 109/L) 0.00-99.99 x109/L
RBC ≤ 0.5% ≤ 1.5% (3.5-6.0 x 1012/L) 0.00-8.00 x1012/L
HGB ≤ 0.5% ≤ 1.5% (110-180g/L) 0-250g/L
PLT ≤ 1.0% ≤ 4.0% (150-500 x 109/L) 0-1000 x109/L
Data Storage Capacity:
Up to 40,000 results including numeric and graphical information
Communication:
LAN Port supports HL7 protocol
Operating Environment
Temperature: 15 °C~30°C
Humidity: 30~85%
Air Pressure: 70~106 kPa
Power Requirement:
A.C.100-240V ≤ 300VA 50/60Hz
Weight: ≤ 65 Kg